Share:


HIOKI IM3536 LCR METER

Quantity: In Stock



LCR METER IM3536

 

General-Purpose LCR Meters with Measurement Frequency from DC, 4 Hz to 8 MHz


Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components. The IM3536 raises the industry standard for general-purpose LCR meters by offering a wide DC and 4Hz to 8MHz (up to 10MHz special order available) testing frequency while delivering 0.05% accuracy, ideal for evaluating the characteristics of next generation electronic components, including power supply inductors thanks to the instrument’s maximum measurement frequency of 8MHz. (up to 10MHz special order available)

Key Features
• DC, 4 Hz to 8 MHz* measurement frequency *Can be customized up to 10 MHz. Please contact your Hioki distributor or subsidiary for more information.
• High-speed measurement of 1 ms (fastest time)
• High-precision measurement of ±0.05% rdg. (representative value)
• Guaranteed accuracy range from 1 mΩ, low-impedance measurement with unmatched repeatability
• DC bias function: Measure under conditions simulating actual use or in accordance with industry standards
• Exceptional specifications and cost-performance for a wide range of applications, from R&D to production line.

 

Model No. (Order Code)

IM3536  
IM3536-01 Special order products up to 10 MHz

FREQUENTLY ASKED QUESTIONS This product is not supplied with measurement probes or test fixtures. Please select and purchase the measurement probe or test fixture options appropriate for your application separately. All probes are constructed with a 1.5D-2V coaxial cable. For an RS-232C connection: A crossover cable for interconnection can be used. You can use the RS-232C CABLE 9637 without hardware flow control.

 

Guarantee3years
 
LAN
 
GPIB
 
RS232C
 
USB2
 
CE
 
Measurement modes LCR (Measurement with single condition), Continuous testing (Continuous measurement under saved conditions)
Measurement parameters Z, Y, θ, X, G, B, Q, Rdc (DC resistance), Rs (ESR), Rp, Ls, Lp, Cs, Cp, D (tanδ), σ, ε
Measurement range 100 mΩ to 100 MΩ, 10 ranges (All parameters are determined according to Z)
Display range Z: 0.00 m to 9.99999 GΩ, Y: 0.000 n to 9.99999 GS, θ: ± (0.000° to 180.000°), Q: ± (0.00 to 9999.99), Rdc: ± (0.00 m to 9.99999 GΩ),
D: ± (0.00000 to 9.99999), Δ%: ± (0.000 % to 999.999 %), or other
Basic accuracy Z ±0.05% rdg. θ: ±0.03° (representative value, Measurable range: 1 mΩ to 200 MΩ)
Measurement frequency 4 Hz to 8 MHz (5 digits setting resolution, minimum resolution 10 mHz)
Measurement signal level [Normal mode: V mode/CV mode]
4 Hz to 1.0000 MHz: 10 mV to 5 Vrms (maximum 50 mArms)
1.0001 MHz to 8 MHz: 10 mV to 1 Vrms (maximum 10mArms)
[Low impedance high accuracy mode: V mode/CV mode]
4 Hz to 1.0000 MHz: 10 mV to 1 Vrms (maximum 100 mArms)
[Normal mode: CC mode]
4 Hz to 1.0000 MHz: 10 μA to 50 mArms (maximum 5 Vrms)
1.0001 MHz to 8 MHz: 10 μA to 10 mArms (maximum 1 Vrms)
[Low impedance high accuracy mode: CC mode]
4 Hz to 1.0000 MHz: 10 μA to 100 mArms (maximum 1 Vrms)
[DC resistance measurement]
Measurement signal level: Fixed at 1 V
DC bias measurement Generating range: DC voltage 0 V to 2.50 V (10 mV resolution)
In low Z high accuracy mode: 0 V to 1 V (10 mV resolution)
Output impedance Normal mode: 100 Ω, Low impedance high accuracy mode: 10 Ω
Display 5.7-inch color TFT with touch panel
Functions Comparator, BIN measurement (10 categories for 2 measurement parameters), Trigger function, Open/short compensation, Contact check, Panel loading/saving, Memory function
Interfaces EXT. I/O( HANDLER) ,USB, USB flash drive, LAN, GP-IB, RS-232C, BCD
Power supply 100 to 240 V AC, 50/60 Hz, 50 VA max.
Dimensions and mass 330 mm (12.99 in) W × 119 mm (4.69 in) H × 230 mm (9.06 in) D, 4.2 kg (148.1 oz)
Accessories Power cord ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1